+ Compact design
+ Innovative effective pulsing technology
+ Low electronic failure statistics
+ Spare parts availability
- High Pump Chamber damage rate
- High HVPS (charger) failure rate
- Calibration sensitivity
- Unconscious, uninformed service abundance
Our company has serviced and repaired for the above-mentioned
models are given.
Pump Chamber renovation - change
Laser Rod Refurbishment
HVPS repair
Simmer PS repair
Fiber Cable Repair
Onsite service
Moreover, the following parts we can provide.
Focus Lens
Window Lens
Flash Lamp
(Including change)
IMPORTANT NOTICE 1
Pump Chamber renovation - as a result of changes in the Head
45 W for output that should take under 700 volts. This
value you should read from the service`s Power meter.
45 W and 700 Volt Pump Chamber cannot provide criteria is
problematic. In this case, you experience the following
problem.
Excessive voltage rise in a short time
Losing energy of the flash lamps in a short time
HVPS damage
High temperature alarm
Not stable shoot with the 15 head
Leaking water from the pump chamber
IMPORTANT NOTICE 2
Should make the re-calibration of your system is compared with
the old value after especially flash lamp, focus lens and
parts changing. Otherwise, your system will work inefficient
. This control provides you to determine whether the part is
original or not.